Investigation of FID signals obtained by inversion recovery technique with and without albumin in H2O-D2O solutions


Author(s): M Zafer Köylü, Mehmet Güven and Burak Akbudak

Abstract: FID signals which are obtained by Inversion Recovery technique with and without albumin in H2O-D2O solutions were investigated. Measurements were carried out using a 400 MHz Avance Bruker 1H-NMR spectrometer. FID signals were obtained from H2O-D2O solutions prepared in specific proportions. Additionally, the effects of pulse parameters on the signals were observed by changing the pulse repetition times. As a result, the Radiation Damping is increased with the increase of H2O-D2O amount. The Radiation Damping is reduced with the addition of albumin. The null point was not observed accurately in the present of Radiation Damping. In addition, to obtain a good signal the pulse delay times should be selected at appropriate intervals and, TR should be taken to 5T1.

Pages: 537-540  |  413 Views  52 Downloads

download (4658KB)

International Journal of Chemical Studies International Journal of Chemical Studies
How to cite this article:
M Zafer Köylü, Mehmet Güven, Burak Akbudak. Investigation of FID signals obtained by inversion recovery technique with and without albumin in H2O-D2O solutions. Int J Chem Stud 2018;6(6):537-540.
 